Testec SI Series TT-SI 9001 Oscilloscope Probe, Active, Differential Type, 25MHz, 1:10, 1:100, BNC Connector
- RS 제품 번호:
- 224-1219
- 제조사 부품 번호:
- SI-9001
- 제조업체:
- Testec
단종
대체품
이 제품은 현재 공급 불가입니다. 아래 제품들은 당사에서 추천드리는 대체품입니다.
- RS 제품 번호:
- 224-1219
- 제조사 부품 번호:
- SI-9001
- 제조업체:
- Testec
제정법과 컴플라이언스
해당 안됨
- COO (Country of Origin):
- TW
제품 세부 사항
TT-SI-9001 Testec Active Differential Probe
The Testec is active differential oscilloscope probe enables a conventional oscilloscope to display and measure in-circuit waveforms that are referenced to high common-mode voltages. It is capable of safely measuring differential voltages up to ±1000V ac and 1400V ac/dc. It supports bandwidths up to 25MHz.
By selecting attenuation ratio 1:10, this oscilloscope probe can measure signals below 70V to get higher resolution and less noise. Or else, set the attenuation to 1:100 when measuring signals up to 700 V dc
By selecting attenuation ratio 1:10, this oscilloscope probe can measure signals below 70V to get higher resolution and less noise. Or else, set the attenuation to 1:100 when measuring signals up to 700 V dc
Features and Benefits
Bandwidth is DC to 25MHz
Attenuation is 1:10/1:100
Rise time is 14ns
Differential mode input voltage is 1:10 ±70V (DC+peak AC) or 70Vrms 1:100 ±700V (DC+peak AC) or 700Vrms
Input impedance is 4Mohm ll5.5pF each side to ground
Common mode voltage range is 1:10 and 1:100 ±700V (DC+peak AC) or 700Vrms
Non-destructive Max input voltage is 1:10 and 1:100 ±1400V (DC+peak AC) or 1000Vrms
CMRR (typical) is -86dB at50Hz and -66dB at 20kHZ
Offset (typical) is <±5mV
Noise (typical) is 0.7mVRMS
4 x AA batteries
Dimensions (LxWxH): 170x63x21 mm
BNC cable length 950 mm
Input leads length 450 mm
Attenuation is 1:10/1:100
Rise time is 14ns
Differential mode input voltage is 1:10 ±70V (DC+peak AC) or 70Vrms 1:100 ±700V (DC+peak AC) or 700Vrms
Input impedance is 4Mohm ll5.5pF each side to ground
Common mode voltage range is 1:10 and 1:100 ±700V (DC+peak AC) or 700Vrms
Non-destructive Max input voltage is 1:10 and 1:100 ±1400V (DC+peak AC) or 1000Vrms
CMRR (typical) is -86dB at50Hz and -66dB at 20kHZ
Offset (typical) is <±5mV
Noise (typical) is 0.7mVRMS
4 x AA batteries
Dimensions (LxWxH): 170x63x21 mm
BNC cable length 950 mm
Input leads length 450 mm
Applications
Oscilloscope probes are used with oscilloscopes to enable the signals on the circuits being tested to be connected to the oscilloscope so that signals can be measured. These oscilloscope probes can be found in:
Performance conversion
Inverters
Motor speed indicators
Switch mode power supplies
Performance conversion
Inverters
Motor speed indicators
Switch mode power supplies
Certifications
CAT III 1000 VAC
CAT III 1400V
CAT III 1400V
사양
속성 | 값 |
---|---|
Probe Type | Active, Differential |
Attenuation | 1:10, 1:100 |
Bandwidth | 25MHz |
Maximum Common Mode Input Voltage | 700V |
Connector Type | BNC |
Model Number p | TT-SI 9001 |
Number of Probes | 2 |
Safety Category Level | CAT III |
Series | SI |
Safety Category | CAT III 1000 V, CAT III 1400V |
Rise Time | 14ns |